expect about 100% RH for the sensors placed over the tap water in a plastic box
Measurement for time 11:10
and 22:50
Average measurement for time 11:10
HIH7120 | SHT85 | AHT10 | AHT20 | AHT21 | AHT25 |
99.99 | 94.65 | 99.50 | 100.00 | 96.58 | 99.97 |
and 22:50
HIH7120 | SHT85 | AHT10 | AHT20 | AHT21 | AHT25 |
100.00 | 99.71 | 100.00 | 100.00 | 99.71 | 100.00 |
Measurement for time 11:10
Measurement for time 22:50
Average measurement for time 21:10
HIH7120 | SHT85 | AHT10 | AHT20 | AHT21 | AHT25 |
28.88 | 28.84 | 28.83 | 28.41 | 28.57 | 29.33 |
Average measurement for time 22:50
HIH7120 | SHT85 | AHT10 | AHT20 | AHT21 | AHT25 |
29.14 | 29.11 | 29.10 | 28.67 | 28.95 | 29.59 |